Rapid Fire Abstracts
Tess E. Wallace, PhD
Senior Clinical MR Research Scientist
Siemens Medical Solutions USA, Inc.
Tess E. Wallace, PhD
Senior Clinical MR Research Scientist
Siemens Medical Solutions USA, Inc.
Patrick Pierce, BSc
MRI Technologist
Beth Israel Deaconess Medical Center
Jennifer Rodriguez
Clinical Trials Specialist
Beth Israel Deaconess Medical Center
Julius Johnson, RT
MRI Technologist
Beth Israel Deaconess Medical Center
Kelvin Chow, PhD
MR Collaboration Scientist
Siemens Healthcare Ltd., Canada, Canada
Xiaoming Bi, PhD
Director, Cardiovascular MR Collaborations
Siemens Medical Solutions USA, Inc.
Reza Nezafat, PhD
Professor
Harvard Medical School
There is a growing need for CMR imaging of patients with cardiac implantable electronic devices (CIED), particularly for assessing scars with late gadolinium enhancement (LGE).1 Although many recent CIEDs are approved for imaging at 3T, CMR imaging for these patients is rarely performed at 3T. Recent data shows that 2/3 of CMR studies in heart failure patients from the SCMR registry were imaged at 3T,2 highlighting the growing need for robust imaging of CIED patients at 3T. In this study, we sought to develop an LGE imaging sequence using a spoiled GRE readout with two echoes, allowing fat-water separation using the Dixon technique to reduce artifacts related to poor fat suppression at 3T in patients with CIED.
Methods:
A 2D ECG-triggered GRE research sequence was implemented with two bipolar echoes and a non-selective adiabatic hyperbolic secant inversion recovery pulse with a spectral bandwidth of 4.0 kHz (Figure 1A). Imaging parameters were as follows: TE1/TE2 1.21/2.64 ms, FA 12°, readout FOV 380 mm, in-plane resolution 1.6 mm, slice thickness 8 mm, bandwidth 801 Hz/px, GRAPPA R=2. Imaging was performed every other heartbeat to allow for signal recovery. Water and fat images were reconstructed using the two-point Dixon algorithm.3
All imaging was performed at 3T (MAGNETOM Vida, Siemens Healthineers, Forchheim, Germany). A phantom comprising doped agarose vials and two oil bottles was scanned with applied B0 field variation to test the effectiveness of fat-water separation in the presence of B0 inhomogeneity. Eight patients (7 males, 66 ± 14 years) with implanted cardiac defibrillators (ICD, N=6) or pacemakers (N=2) were scanned following informed consent. Patients were positioned with arms above their heads and imaged during end-inspiration to maximize the distance between the device and the region of interest. LGE imaging was performed 15-20 minutes after injection of 0.1 mL/kg Gadovist (Bayer, Berlin, Germany). Optimal inversion times (TI) were determined using a free-breathing Look-Locker sequence.
Results:
Water-only and fat-only images obtained in a phantom with applied B0 field variations are shown in Figure 1B. Dixon fat-water separation is robust in the presence of a large applied magnetic field gradient, and the vial representing a healthy myocardium is successfully nulled. Fat-water separation in the heart was successful in all subjects. Figure 2 shows example water-only and fat-only images in multiple cardiac views in a patient with ICD, where robust water-fat separation is achieved. Examples of myocardial scar detection in patients with ICDs are shown in Figure 3. On in-phase images, it is challenging to differentiate scar tissue from fat, which also appears bright. Regions of LGE are visible in the water-only images.
Conclusion:
LGE-Dixon enables robust fat suppression for LGE imaging in patients with CIED at 3T. Future studies will compare image quality with conventional methods in a larger patient population.